F950 core specifications
Detection Sensitivity Quinine sulfate solution detection: 1×10⁻⁹ g/mL
Excitation Wavelength Range 360–600nm (LED)
Emission Wavelength Range 360–650nm (extendable to 900 nm)
Emission Bandwidth 12nm
Emission Wavelength Accuracy ±2.0nm
Emission Wavelength Repeatability ≤1.0nm
Peak Intensity Repeatability ≤1.5%
Measurement Linearity Correlation coefficient ≥ 0.995
Fluorescence Detector Original Hamamatsu photomultiplier tube
Data Transmission Mode RS232 serial port, USB-B, USB-A
PC Software Optional (optional 21 CFR compliant software)
Power 220V/50HZ

Send your sample type and method, and we will confirm the configuration and current lead time.

Ask about this model

[email protected]

Overview

What this instrument measures, and how it fits the range.

F950 is a fluorescence spectrophotometer in the F Series, with an excitation wavelength range of 360–600 nm (LED) and an emission wavelength accuracy of ±2.0 nm.

Key points

  • Manual wavelength adjustment reduces the need for a motorized drive system. The instrument supports batch quantification at fixed wavelengths.
  • The compact optical system uses few moving parts, an LED light source and a photomultiplier tube.
  • Optional excitation LED sources: 365nm, 376nm, 392nm, 405nm.

Full specifications for this model are listed at the top of this page.

Accessories

Confirm these items before the final quotation. Accessories are listed only where the manufacturer offers them.

Standard accessories

Ask for an itemized quotation that separates standard items from options.

Optional accessories

  • Excitation LED sources: 365 nm, 376 nm, 392 nm, 405 nm
  • PC software
  • 21 CFR compliant software
Specifications can depend on the selected configuration. Confirm the current datasheet, accessories and software version before purchase.

Applications for this instrument

See how the model is positioned for real measurement tasks.

Let’s talk about your application

Find the right instrument for your application

Share your sample, measurement requirements and preferred configuration.

Contact our team