Explore ApplicationsKE Series
Volumetric Karl Fischer titrator measuring 0.1-250 mg of water with 0.1 mg resolution, reporting results in mg, mg/L, percent and ppm.
Documentation
Datasheets and manuals are issued per model and revision. Ask for the documents that match the configuration you are considering.
See linked documents| Measurement Range | 0.1~250mg |
|---|---|
| Resolution | 0.1mg |
| Measurement Result | mg, mg/L, %, ppm |
| Polarization Current | 1μA±0.2μA, 50μA±10μA |
| Indication Error | ≤5% |
| Repeatability | ≤0.5% |
| Accounts | 100 pcs |
| Methods | 100 sets |
| Data Storage | 10000 sets |
| Output | USB port for data export |
| PC Software | Not available |
| Printing | Optional wired dot-matrix printer and Bluetooth thermal printer. |
Send your sample type and method, and we will confirm the configuration and current lead time.
Ask about this modelWhat this instrument measures, and how it fits the range.
KE-01 is a Karl Fischer volumetric titrator in the KE Series, with a measurement range of 0.1~250 mg and a resolution of 0.1 mg.
Full specifications for this model are listed at the top of this page.
Confirm these items before the final quotation. Accessories are listed only where the manufacturer offers them.
Ask for an itemized quotation that separates standard items from options.
Confirm requirements with our team before the final order.
The range is 0.1–250 mg of water with 0.1 mg resolution, reported in mg, mg/L, percent or ppm. The practical ppm limit depends on sample size: with a 5 g sample, 0.1 mg of water corresponds to about 20 ppm.
Volumetric titration suits samples with appreciable water content — solvents, oils, food and pharmaceutical excipients in the 0.1–250 mg range. Coulometric titration is the better fit at trace levels below roughly 50 ppm. The KE-01 is a volumetric instrument.
100 methods, 100 user accounts and 10,000 stored data sets, with USB export. The specified repeatability is ≤0.5% and the indication error ≤5% — quote those two figures when you validate the method.